18 Mobius Tutorial: Cahper 2 - fEPSPs Recording
2-1. Recording of fEPSPs
2-1.3. Running workflow
The workflow is run and stopped (paused) with the execution buttons.
Fig.2-1.6. Execution buttons.
2-1.4. Before starting your experiment
When a recording workflow is run with Green or Green-Red button for the first time after it is opened
(or made), Mobius needs several seconds for its calibration (as seen in the pop-up message). Acquisi-
tion starts automatically after the calibration.
If you would like to start acquisition immediately after clicking the Green-Red button, run the workflow
with the green button for a few traces with both stimulators un-enabled, and then stop before starting
your experiment. When the workflow is run the next time, Mobius does not require calibration but
acquisition will start immediately.
Stop Run WITHOUT Run WITH Pause
saving data saving data
Fig.2-1.7. The top shows pop-up mesage for the calibration. Acquisition
always starts after this message disappears.
In order to start recording immediately after clicking the Green-Red button,
run Mobius with the Green button for a few traces with stimulator un-enabled
before starting your experiment in advance. (bottom figure.)
Comments to this Manuals