Sub-menu Settings
step on
Sinking Time
The sinking time permits a controlled return of the machine's piston after the break, i.e. after this time
the piston is stopped. The sinking time is activated by the specimen's break or by the stop key. With
stop the sinking time can be aborted. After this time it can be activated by stop once again, that's why
the piston is sinked completely.
Automatic Zeroizing
1-2 seconds after starting the display is automatically set to zero.His important that there is not yet
a frictional connection, as otherwise the breaking load won't be correct.
For this zeroizing the machine needs approx. 2 seconds. So please take care that the sinking time is
respectively set, so that with the next test after the start signal there are at least 2 seconds left for the
piston's extension to the specimen.
Displacement Measuring: (Option)
AKT: travelling displacement display, which always measures and indicates the actual displacement.
HLD: displacement display is stopped at force maximum, i. e. it is only measured in case of increasing
force. PEAK-measurement must be on.
Peak value
Force
Displacement Hold
MAX: displacement display is stopped at maximum value, i.e. the maximal displacement is measured
and indicated. PEAK-measurement must be on.
Set up
M1
Sink time: ON
: O
N
Disp.measur: ACT
%
: Choice
: Abort
Time: 3.0s
Auto Zero
Pacemaker:
FO
Zeroband: ON
Break detec: OFF
Level: 20.0
MA
ESC